Relative accuracy of bioelectrical impedance analysis for assessing body composition in children with severe obesity

Principle/Accuracy
Author
Khan et al.
check :

Used Parameters : PBF, BFM, ASM

Publication Year : 2020

Country : USA

Journal : Journal of Pediatric Gastroenterology and Nutrition

Objectives: The accuracy of different bioelectrical impedance analysis (BIA) devices for assessing body composition in children with obesity is unclear. We determined the relative accuracy of 2 BIA devices compared to dual x-ray absorptiometry (DXA) in obese and severely obese children. Methods: We measured body composition in a cross-sectional study of 78 obese children by a handheld single frequency tetrapolar BIA device (Omron), a stationary multifrequency octopolar BIA device (InBody 370) and DXA. Intermethod …
Full text